Model Number 0168L12
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/17/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that it was difficult to inflate the balloon of the catheter with water during pretest.Another catheter was used for the patient.
 
Additional Manufacturer Narrative
Received 1 coude catheter.The reported event was unconfirmed.The sample was visually inspected and did not find anything to contribute to the reported event.Per the functional evaluation, the balloon was inflated with 10cc water and found that the balloon inflated easily.The reported event was unable to be duplicated.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "visually inspect the product for any imperfections or surface deterioration prior to use." (b)(4).
 
Event or Problem Description
It was reported that it was difficult to inflate the balloon of the catheter with water during pretest.Another catheter was used for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER COUDE TIP
Common Device Name
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key6880272
Report Number1018233-2017-04909
Device Sequence Number10967731
Product Code EZC
UDI-Device Identifier00801741017803
UDI-Public(01)00801741017803
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2021
Device Model Number0168L12
Device Catalogue Number0168L12
Device Lot NumberPK6238
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2017
Initial Date Received by Manufacturer 08/28/2017
Supplement Date Received by Manufacturer11/09/2017
Initial Report FDA Received Date09/20/2017
Supplement Report FDA Received Date11/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-