Model Number H7493926016250
Device Problem Hole In Material (1293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/27/2017
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.(b)(4).Device evaluated by mfr.: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental med watch will be filed.(b)(4).
 
Event Description
It was reported that balloon pinhole occurred.The 80% stenosed target lesion was located in a moderately tortuous and severely calcified coronary artery.A 2.50 x 16 synergy ii drug-eluting stent was advanced and difficulties were encountered crossing the lesion.During the first inflation of the stent delivery balloon at 16 atmospheres the balloon did not inflate and a pinhole was noted.The procedure was completed with a different device.No patient complications were reported and the patient's status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Type of Device
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6880948
MDR Text Key87885323
Report Number2134265-2017-09689
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08714729840091
UDI-Public08714729840091
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/30/2018
Device Model NumberH7493926016250
Device Catalogue Number39260-1625
Device Lot Number20260706
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/31/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-