| Model Number |
ZXR00 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).An attempt has been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxr00 26.5 diopter intraocular lens (iol) was explanted on (b)(6) 2017 due to the post-operative myopic outcome.The lens was replaced with the same model, but different diopter (25.0).No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 9/20/2017.Device returned to manufacturer? yes.Device evaluation: visual inspection of the return sample shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the returned lens, additional analysis is not possible.The customer's reported complaint cannot be confirmed.Manufacturing record review: manufacturing records were reviewed and the device was manufactured according to specification.A search on complaints history revealed no similar complaints were received for this production order number to date.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
|
| |
|