Pump passed operating current, displacement test but alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the occlusion, prime and excessive no delivery test due to a33 alarm.No moisture damage on the lcd board, mother board, interface board, rf board, motor, vibrator motor and battery tube assembly during visual inspection.Pump received with cracked case at the display window corner, cracked reservoir tube lip, minor scratched lcd window, stained end cap sticker, stained address/serial number label, cracked belt clip slot and cracked battery tube threads.
|