The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of gfav3665 shows eleven other similar product complaint(s) from this lot number.The complaints for this lot number (gfav3665) have been reported from the same facility.
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a damaged stylet was confirmed but the exact cause was unknown.Ten navarre drainage catheter kits were returned for investigation.One kit was open and nine kits were sealed.Contained in the open kit (sample #1) was a broken flexible stylet.The injection cap was attached to the stylet luer adaptor.There was a break in the stylet at the distal end of the white connector and 1.3cm from the distal end of the white connector.The stylet was kinked 1.1cm distal to the break.The sealed packages were labeled samples #2 - #10.The stylet for sample #2 was kinked approximately 2cm and 4cm distal to the white connector.Sample #3 exhibited a kink 1.4cm from the white connector and a break 2.5cm from the white connector.Sample #4 exhibited a break 1.5cm from the distal end of the white connector.Sample #5 exhibited a break 1.0cm from the distal end of the white connector.No breaks were observed on samples #6, #7, #8, #9, & #10.Sample #8 was opened for additional investigation.The clear portion of the stylet extended approximately 18.5cm from the distal end of the label and approximately 8.7cm from the proximal end of the label.The stylet was removed and flexed 4.5cm, 18.5cm, & 41.5cm from the distal end of the white connector.The stylet snapped and fractured at each flex point.A microscopic examination of the fractured ends of the stylet revealed jagged edges.The complaint samples were forwarded to the manufacturing facility for further review.A review of the manufacturing records showed no potential contributing factors to the stylet damage.The product ifu cautions, ¿do not use excessive force when inserting stiffening cannula to straighten catheter (pigtail).Catheter tip may be straightened manually to aid in insertion of stiffening cannula.¿ a lot history review (lhr) of gfav3665 shows eleven other similar product complaint(s) from this lot number.The complaints for this lot number (gfav3665) have been reported from the same facility.
|