| Catalog Number |
9734403 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/25/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect thoracic probe has been received by manufacturer for evaluation.Testing found that the tip of the probe had nicks and was bent to one side.The back end of the probe had several impact marks not allowing the probe to insert into the mating navlock tracker.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that during a spinal fusion procedure while in navigation when using the thoracic probe to start entry point in the bone the tip of the probe bent.The surgeon completed the procedure with another probe.The procedure was completed with the use of navigation.There was a delay of less than 1 hour.No impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: unique device identification (udi) updated to proper value.
|
| |
|