| Model Number |
7050-90 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Undesired Nerve Stimulation (1980)
|
| Date of Event |
09/08/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the information provided, review of the surgical technique manual, ifu, certificates of conformance and fmea, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is placing the implant too deep possibly due to the patient's abnormal anatomy.
|
| |
|
Event or Problem Description
|
|
The patient had left side si joint arthrodesis in (b)(6) 2017 where two implants were placed.The patient had si joint related pain relief, but complained of leg pain following the initial procedure.The surgeon determined that the first implant was impinging on a nerve.In (b)(6) 2017, the surgeon performed a revision surgery where he backed out the first implant approximately one centimeter to alleviate the nerve impingement.The patient's leg pain resolved following the revision procedure.
|
| |
|