Model Number 7050-90
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Undesired Nerve Stimulation (1980)
Date of Event 09/08/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, review of the surgical technique manual, ifu, certificates of conformance and fmea, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is placing the implant too deep possibly due to the patient's abnormal anatomy.
 
Event or Problem Description
The patient had left side si joint arthrodesis in (b)(6) 2017 where two implants were placed.The patient had si joint related pain relief, but complained of leg pain following the initial procedure.The surgeon determined that the first implant was impinging on a nerve.In (b)(6) 2017, the surgeon performed a revision surgery where he backed out the first implant approximately one centimeter to alleviate the nerve impingement.The patient's leg pain resolved following the revision procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
3055 olin avenue
suite 2200
san jose CA 95128 2066
MDR Report Key6888138
Report Number3007700286-2017-00092
Device Sequence Number705941
Product Code OUR
UDI-Device Identifier00859256003309
UDI-Public00859256003309
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/19/2021
Device Model Number7050-90
Device Lot Number493753
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2017
Initial Report FDA Received Date09/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
-
-