| Catalog Number |
04P75-31 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, abbott point of care (apoc) was contacted by a customer who reported visible smoke and a burning smell emanated from analyzer (b)(4).There was no additional information at the time of this report.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.However, the customer does not state if 9 volt or rechargeable batteries were being used.Therefore the analyzer is unlikely to become hot to touch.The product was replaced and returned for investigation.Based on the information available, there were no patient or user related injuries associated with this complaint.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The investigation was completed on 11/02/2017.The failure was due to a defective tantalum capacitor.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|