MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY Back to Search Results
Catalog Number 04P75-31
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2017, abbott point of care (apoc) was contacted by a customer who reported visible smoke and a burning smell emanated from analyzer (b)(4).There was no additional information at the time of this report.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.However, the customer does not state if 9 volt or rechargeable batteries were being used.Therefore the analyzer is unlikely to become hot to touch.The product was replaced and returned for investigation.Based on the information available, there were no patient or user related injuries associated with this complaint.
 
Additional Manufacturer Narrative
(b)(4).The investigation was completed on 11/02/2017.The failure was due to a defective tantalum capacitor.
 
Event or Problem Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT1 ANALYZER, IMMUNO READY
Common Device Name
I-STAT1 ANALYZER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key6888239
Report Number2245578-2017-00309
Device Sequence Number820160
Product Code CGA
UDI-Device Identifier00054749003709
UDI-Public00054749003709
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K001387
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04P75-31
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2017
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/18/2017
Supplement Date Received by Manufacturer11/02/2017
Initial Report FDA Received Date09/22/2017
Supplement Report FDA Received Date11/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-