| Model Number |
388.509 |
| Medical Device Problem Codes |
Break (1069); Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/02/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.(b)(6).The investigation could not be completed; no conclusion could be drawn, as no product was received.Dhr review for part# 388.509 lot# a7pa03 synthes lot # 5164062.Release to warehouse date: 30jan2006, supplier: (b)(4), no ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes (b)(4) reports an event in (b)(6) as follows: it was reported that during a lumbar discectomy and fusion on (b)(6) 2017 using the uss (universal spinal system), the rod introduction pliers device was being used to seat the uss collar over the screw and tighten onto the rod.This device has a collet that slides up and down and the collet is gently tapped in order to bring down a "sleeve cap" (uss collar) into position.As the surgeon impacted it, this collet (sliding portion) broke off.The part was manually retrieved and surgery was completed with a second similar device.There was no surgical delay, and no reported harm to the patient.This complaint involves one device.This report is for one (1) rod introduction pliers.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Subject device has been received and is currently in the evaluation process.Investigation is ongoing; no conclusion could be drawn as product is entering the complaint system.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product development investigation was completed.The report indicates that the: the rod introduction pliers are a component of the dual-opening uss system and are utilized for construct assembly.A collar can be loaded onto the barrel of the rod introduction pliers.The pliers¿ jaws can be utilized to engage the rod and the handles compressed to insert the rod into the hook or screw.Finally the collar can be seated on the screw using a hook positioner (388.631) and a 12-point nut attached and tightened.The returned instrument was evaluated and the complaint condition was able to be confirmed as the collet was found to be broken at the silver braze.The remainder of the device was found to function as intended.The complaint condition was unable to be replicated due to post-manufacturing damage.No dimensional analysis was available at the fracture site due to post-manufacturing damage.Based on the complaint description the collet was impacted causing the device to fail.Relevant drawings for the returned instrument were reviewed (both current revision and from the time of manufacture: top-level.The design, materials and finishing processes were found to be appropriate for the intended use of this device.A device history review, including material and hardness review, was performed for the returned instrument¿s lot number and no ncrs or complaint-related issues were identified with the lot number which may have contributed to the complaint condition.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.Device used for treatment, not diagnosis.Report was initially submitted on oct 27, 2017, but the fda site was down.Advised by fda on nov 8, 2017 to resubmit medwatch.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|