| Catalog Number |
RBY4C0202 |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/24/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2017-01650.The hospital disposed of the device.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure to treat a bleed in a branch of the superior femoral artery (sfa) using ruby coils and a ruby coil detachment handle (handle).During the procedure, the physician deployed a ruby coil into the target vessel using a lantern delivery microcatheter (lantern) and then attempted to detach the coil using the handle.However, after multiple attempts, the ruby coil was unable to detach.Therefore, the ruby coil was removed and the procedure was completed using additional ruby coils and the same handle.There was no report of an adverse effect to the patient.
|
| |
|