• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY4C0202
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/24/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2017-01650.The hospital disposed of the device.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a bleed in a branch of the superior femoral artery (sfa) using ruby coils and a ruby coil detachment handle (handle).During the procedure, the physician deployed a ruby coil into the target vessel using a lantern delivery microcatheter (lantern) and then attempted to detach the coil using the handle.However, after multiple attempts, the ruby coil was unable to detach.Therefore, the ruby coil was removed and the procedure was completed using additional ruby coils and the same handle.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6888993
Report Number3005168196-2017-01649
Device Sequence Number10921795
Product Code HCG
UDI-Device Identifier00814548013121
UDI-Public00814548013121
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/02/2024
Device Catalogue NumberRBY4C0202
Device Lot NumberF71897
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2017
Initial Report FDA Received Date09/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
-
-