• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT300
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (model zxt300, +14.0 diopter) was explanted in a secondary surgical procedure from the left eye of a female patient because the lens was observed rotated post implantation.Reportedly, another lens (model zxt225, +13.5 diopter) was implanted as a replacement without incidents.Additional information was received and it was learnt that the lens had rotated about 20 degrees post implantation.The surgeon decided to perform a lens exchange instead of a lens rotation.No specific patient issues were reported.No incision enlargement, no vitrectomy was performed and no sutures were used.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation?: yes, returned to manufacturer on 09/14/2017.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6889115
Report Number9614546-2017-00914
Device Sequence Number10926784
Product Code POE
UDI-Device Identifier05050474606265
UDI-Public(01)05050474606265(17)211024
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/24/2021
Device Model NumberZXT300
Device Catalogue NumberZXT300U140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2017
Initial Date Received by Manufacturer 08/30/2017
Supplement Date Received by Manufacturer10/20/2017
Initial Report FDA Received Date09/23/2017
Supplement Report FDA Received Date11/15/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-