Model Number ETLW1610C93E
Medical Device Problem Code Occlusion Within Device (1423)
Health Effect - Clinical Codes Occlusion (1984); Pain (1994)
Date of Event 08/25/2017
Type of Reportable Event Serious Injury
Event or Problem Description
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of a 6 cm in diameter abdominal aortic aneurysm.It was reported that the patient presented emergently with leg pain and a spiked creatinine level.A ct scan was preformed and revealed that the left endurant iliac limb and left external iliac artery were occluded, due to tortuous anatomy.The physician performed a thrombectomy with the use of an angiojet machine and the ekos machine to dissolve the clot.After the iliac limb was patent a 10x 40 bare metal stent was implanted in the distal end of the endurant stent graft.Per the physician, the cause of the occlusion was related to the patient¿s anatomy, the distal end of the endurant stent graft was implanted near a kink of the patient¿s native external iliac artery.No additional clinical sequelae were reported and the patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II ILIAC STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key6890316
Report Number2953200-2017-01481
Device Sequence Number655742
Product Code MIH
UDI-Device Identifier00643169467606
UDI-Public00643169467606
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/29/2019
Device Model NumberETLW1610C93E
Device Catalogue NumberETLW1610C93E
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2017
Initial Report FDA Received Date09/25/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
-
-