LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Electrical Power Problem (2925)
Patient Problem No Patient Involvement (2645)
Event Date 08/30/2017
Event Type  Malfunction  
Manufacturer Narrative

There was no patient involvement. Contact telephone number:(b)(6). The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601). Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). This medwatch report is being filed on behalf of livanova (b)(4). A livanova field service representative was dispatched to the facility to investigate. The service representative confirmed the reported issue and identified a faulty compressor. The customer decided not to proceed with a repair. The device was isolated and the customer plans to scrap the device due to age. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. Device was scrapped.

 
Event Description

Livanova (b)(4) received a report that upon startup of the heater-cooler system 3t, the mains power shut down as the compressor started. There was no patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6892361
MDR Text Key87786005
Report Number9611109-2017-00753
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation HEALTH PROFESSIONAL
Remedial Action Other
Type of Report Initial
Report Date 09/25/2017
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received09/25/2017
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-80
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/31/2017
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured07/17/2007
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-