• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 363 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 363 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 363LF-410
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/15/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Device labeling: potential adverse events that may occur with saline-filled breast implant surgery include: reoperation, pain, wrinkling, asymmetry, implant palpability/visibility, implant removal, capsular contracture, changes in nipple and breast sensation, implant displacement/migration, implant deflation, scarring, infection, hematoma/seroma, breastfeeding complications, implant extrusion, necrosis, delayed wound healing, breast tissue atrophy/chest wall deformity, calcium deposits, and lymphadenopathy.Deflation ¿ breast implants are not lifetime devices.Saline breast implants deflate when the shell develops a tear or hole.Deflation can occur at any time after implantation, but they are more likely to occur the longer the implant is implanted.The following things may cause implants to deflate: damage by surgical instruments; folding or wrinkling of the implant shell; excessive force to the chest (e.G., during closed capsulotomy, which is contraindicated); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.Laboratory studies have identified some of the causes of deflation for allergan¿s product; however, it is not conclusively known whether these tests have identified all causes of deflation.
 
Event or Problem Description
Patient reported a right side deflation.Device remains implanted.
 
Event or Problem Description
Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: one curved opening in the valve bond edge and white particles in inner surface.A leak test was performed which identified two openings on the device.Microscopic analysis was performed which identified: one curved opening in the valve bond edge displayed sharp edge opening (unidentified (tear) opening) and one curved opening in the posterior side displayed sharp edge opening (unidentified (tear) opening).A dimension measurement in the shell was performed which identified the thickness within specification.A fill inspection was performed and identified no blockage in valve.Based on the device analysis the final assessment was: two unidentified (tear) openings.
 
Event or Problem Description
Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 363 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key6894270
Report Number9617229-2017-02625
Device Sequence Number699450
Product Code FWM
UDI-Device Identifier10888628001022
UDI-Public(01)10888628001022(10)2683845(11)141217(17)181117
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/17/2018
Device Model Number363LF-410
Device Catalogue Number363LF-410
Device Lot Number2683845
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2018
Initial Date Received by Manufacturer 09/07/2017
Supplement Date Received by Manufacturer01/31/2018
03/01/2018
Initial Report FDA Received Date09/26/2017
Supplement Report FDA Received Date02/22/2018
03/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-