MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM10040
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/05/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or no medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation to date.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an endovascular stent graft procedure to seal an aneurysm in the left arm transposed av fistula; basilica vein, the stent graft allegedly jumped and was implanted approximately 2cm from the intended placement site.Reportedly, the delivery system was removed without incident, and a second endovascular stent graft was deployed to completely seal the aneurysm.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing related cause was considered.The device history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.The review showed no remarkable incidents occurred during the manufacturing process.No additional complaint has been previously reported for this lot number.Investigation summary: as part of all complaint investigations, an attempt was made to evaluate the subject device as well as the device usage.In this particular case no sample was provided for evaluation.Based on the available information and as no sample was returned the investigation was closed with inconclusive results.A definite root cause for the reported event could not be determined.However, the reported application represents an off label use of the device.Labeling review: in reviewing the labeling supplied with this product it was found that the instructions for use (ifu) sufficiently address the potential risks.The ifu states regarding increased deployment forces: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device of the same size." regarding the anatomy of the placement site the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." furthermore, the ifu states: "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment.¿ the reported application represents an off label use of the device.The ifu states that the safety and effectiveness of the device when placed across an aneurysm or a pseudoaneurysm has not been evaluated.
 
Event or Problem Description
It was reported that during an endovascular stent graft procedure to seal an aneurysm in the left arm transposed av fistula; basilica vein, the stent graft allegedly jumped and was implanted approximately 2cm from the intended placement site.Reportedly, the delivery system was removed without incident, and a second endovascular stent graft was deployed to completely seal the aneurysm.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6895022
Report Number2020394-2017-01224
Device Sequence Number761699
Product Code PFV
UDI-Device Identifier04049519008585
UDI-Public(01)04049519008585
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2020
Device Model NumberFEM10040
Device Catalogue NumberFEM10040
Device Lot NumberANBQ4707
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/30/2017
Supplement Date Received by Manufacturer11/21/2017
Initial Report FDA Received Date09/26/2017
Supplement Report FDA Received Date11/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
Patient Weight75
-
-