• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION E SERIES DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION E SERIES DEFIBRILLATOR Back to Search Results
Model Number E SERIES
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/05/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that during a routine shift check by a clinician, the device failed self-test.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation evaluated the device and the reported malfunction was not replicated or confirmed.The device was put through extensive testing without duplicating the malfunction.The device was recertified and returned to the customer.Review of the activity logs did not find evidence to support the reported event.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E SERIES DEFIBRILLATOR
Common Device Name
E SERIES
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key6895061
Report Number1220908-2017-02301
Device Sequence Number19034931
Product Code MKJ
UDI-Device Identifier00847946009786
UDI-Public00847946009786
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K042007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberE SERIES
Device Catalogue NumberE SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2017
Initial Date Received by Manufacturer 09/05/2017
Supplement Date Received by Manufacturer09/05/2017
Initial Report FDA Received Date09/26/2017
Supplement Report FDA Received Date10/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-