• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GALIL MEDICAL LTD ICESPHERE 1.5 90 CRYOABLATION NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GALIL MEDICAL LTD ICESPHERE 1.5 90 CRYOABLATION NEEDLE Back to Search Results
Model Number FPRPR3560
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/29/2017
Type of Reportable Event Malfunction
Event or Problem Description
Prior to a cryoablation procedure, two icesphere 1.5 90 needles and system tested fine with no issues to report.Towards the end of the first freeze cycle the doctor noticed the shaft of one of the needle started to collect frost.The patient's skin was covered with saline to prevent any injury.The procedure was completed as expected with no injury to the patient.The size of the iceball was reportedly good.
 
Additional Manufacturer Narrative
The needle was returned for investigation.The manufacturing records were reviewed and no related deviation or concerns were found.The reported frost on the needle shaft was confirmed as the needle failed investigative testing.The needle was disassembled and inspected.No visible soldering gaps or voids were found.Microscopic inspection of the vacuum sleeve was conducted and no corrosive pits or soldering flux residuals were identified on the external surface of the vacuum sleeve.A bubble test of the vacuum sleeve was conducted; no leaks were identified.Based on the investigation results, the customer complaint was verified and the root cause was determined to be a component failure.No relationship was identified between the needle assembly processes and the vacuum loss phenomena.The treatment was completed with no injury to the patient as the physician used a warm saline and a glove to protect the patient's skin.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICESPHERE 1.5 90 CRYOABLATION NEEDLE
Common Device Name
CRYOABLATION NEEDLE
Manufacturer (Section D)
GALIL MEDICAL LTD
1 tavor building, p.o. box 224
industrial park
yokneam,, israel 20692
IS  20692
MDR Report Key6895469
Report Number9616793-2017-00110
Device Sequence Number636785
Product Code GEH
UDI-Device Identifier07290104830945
UDI-Public07290104830945
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K152133
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date12/01/2017
Device Model NumberFPRPR3560
Device Lot NumberB8326-038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2017
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/31/2017
Supplement Date Received by Manufacturer10/01/2017
Initial Report FDA Received Date09/26/2017
Supplement Report FDA Received Date10/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-