CYBERONICS, INC. LEAD MODEL 304 Back to Search Results
Model Number 304-20
Device Problem High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/01/2017
Event Type  malfunction  
Event Description
It was reported that the patient was in clinic for the first appointment around 2 months after vns implantation and high impedance was observed on the patient's device.The patient was scheduled for a surgery to address the issue.No surgical intervention is known to have occurred to date.No additional relevant information has been received to date.
 
Event Description
Additional information was received that a revision surgery took place due to the high impedance and it was found that the lead was not fully inserted into the generator which was causing the high impedance.The pin was re-inserted into the generator and the impedance value was found to be ok.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 304
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6896389
MDR Text Key87489952
Report Number1644487-2017-04511
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/27/2020
Device Model Number304-20
Device Lot Number203869
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received10/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age16 YR
-
-