| Model Number |
HEWROOT0030 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/29/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) the product was returned for evaluation.A mismatch between the graft itself (axillo bifemoral knitted graft, radially supported, diameter 8 mm, length 60 cm body/30 cm branch) and its labeling (cardioroot graft, diameter 30 mm) was confirmed.A non conforming report has been initiated.The most probable identified root cause is a product mix-up occuring during a manufacturing step.
|
| |
|
Event or Problem Description
|
|
It was reported that before a surgery, while opening the packaging, the device found in the box was significantly different than the labelled one.The device has not been used has been returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The conducted investigation concludes that the root cause is a product mix-up between two products.The inversed device is located in mexico and will be returned as part of a recall.No other product was involved in the reported event.
|
| |
|