BAYER MEDICAL CARE, INC. STELLANT CT SYRINGE KIT; ANGIOGRAPHIC SYRINGE KIT Back to Search Results
Model Number 60728737
Medical Device Problem Codes Product Quality Problem (1506); Device Contaminated During Manufacture or Shipping (2969); Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/31/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Bayer quality assurance product analysis received and examined the returned low pressure connecting tubing (lpct), lot #8532106.The defect was identified as degraded compound (resin) that is partially embedded in the tubing wall and extends into the fluid pathway.The cause of the reported problem is degraded compound (resin) that collected on the extrusion tooling and became partially embedded in the tubing as it was formed during manufacturing.A 12 month review of complaint history determined the frequency to be (b)(4) complaints per million (cpm) for similar defects.
 
Event or Problem Description
The site reported the following: a stellant disposable set has visible debris in the line.This is a sds-ctp-spk syringe kit.The kit was removed from the area and was not used for a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLANT CT SYRINGE KIT
Common Device Name
ANGIOGRAPHIC SYRINGE KIT
Manufacturer (Section D)
BAYER MEDICAL CARE, INC.
one bayer drive
indianola PA 15051
MDR Report Key6896838
Report Number2520313-2017-00062
Device Sequence Number11114793
Product Code DXT
UDI-Device Identifier00616258005188
UDI-Public(01)00616258005188(11)170626(17)210626
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K082905
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 09/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/26/2021
Device Model Number60728737
Device Catalogue NumberSDS-CTP-SPK
Device Lot Number8532106
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/31/2017
Initial Report FDA Received Date09/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-