| Model Number |
201106 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complaint device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the complaint device, a bend was identified approximately 4 centimeters from the distal tip.Device inspection revealed the catheter was unable to meet specification for curve and planarity.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The most likely root cause is bend in the catheter shaft resulting in a curve and planarity issue as a result of mishandling subsequent to distribution, including shipping/storage conditions or improper manipulation of the catheter.The instructions for use (ifu) state: - do not introduce the tip folded into the guiding sheath.- do not exert excessive pressure during placement of catheter if unknown resistance is encountered.- do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.- avoid manual pre-bending of distal curve, as this may damage steering mechanism of steerable catheters.The reported event will continue to be monitored through post-market surveillance.
|
| |
|
Event or Problem Description
|
|
It was reported that the shape of the electrophysiology catheter was not correct.Multiple attempts were made to obtain additional information.No information was provided.However, there was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received from the customer after filing the initial mdr.The executive summary was updated to include this information.The reported event will continue to be monitored through post-market surveillance.
|
| |
|
Event or Problem Description
|
|
It was reported that the shape of the electrophysiology catheter was not correct.The catheter was replaced and the procedure was completed.The complainant was unaware of any patient injury, medical intervention, or extended procedure time.These are commonly used devices that are readily available.
|
| |
|