| Model Number |
MICL12.6 |
| Medical Device Problem Codes |
Tear, Rip or Hole in Device Packaging (2385); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/30/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: lens was returned in liquid in the lens vial.Visual inspection found a piece of the haptic missing.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportedly, as the surgeon went to open the package of a sterile 12.6mm micl12.6 lens, he noticed the tyvek covering was already slightly opened.Surgeon continued to implant the lens.During implant into the right (od) eye, the lens tore.He then removed the lens and used the back-up lens.This occurred on (b)(6) 2017.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: a lens work order search was performed.No similar complaint type events were found.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).
|
| |
|