| Model Number |
505003 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.The unit has been repaired, and recertified per procedure.The unit was restored to proper operation.
|
| |
|
Event or Problem Description
|
|
The unit was sent to a service center for repair.Upon triage, a technician found a burned track on the keyboard.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record shows that this unit was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, corrective actions will be taken as necessary.Very little current is required for the keypad overlay to function so it is designed accordingly.For multiple traces to have ¿burned¿, there is either an overvoltage condition or a chemical/liquid ingress occurrence.If information is provided in the future, a supplemental report will be issued.
|
| |
|