• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE INTEGRITY RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE INTEGRITY RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RSINT22518UX
Medical Device Problem Codes Difficult to Open or Remove Packaging Material (2922); Material Distortion (2977)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/21/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis photo review: a photo provided by the account captured the damage to the distal end of the protective sheath.Product analysis summary: the sheath and stylette were returned loaded in the device.The stent was positioned on the balloon between the marker bands as per specifications.There was no deformation evident to the stent.There was no damage or material break evident to the distal tip.There was no damaged evident to the delivery system catheter.The sheath and stylette were removed from the device with no resistance noted.The tapered section of the sheath was severely bunched and deformed.There was no kinks or damage evident to the stylette.(b)(4).
 
Event or Problem Description
Physician intended to use a resolute integrity rx drug eluting stent.It was reported that there were no issues noted to the packaging, there were no issues noted when removing the device from the hoop/tray.The device was inspected and no issues were noted; no defects in appearance apparent prior to attempting the removal of the stylette or protective sheath.Resistance was noted when attempting to remove the protective stylette and it was stuck.The tech then tried a second time prior to tearing their glove.Their glove became snagged on the stylette and was stuck on the open area of the stylet loop and the protective sheath was resistant to removal and became misshapen.A break occurred at the tip.Negative prep was not performed.The device was not kinked and re straightened during use.Neither the tech or physician attempted to remove the protective sheath once the device had been contaminated.The device was not used in the patient as the torn glove contaminated the product so the team removed it from the table.The device was replaced with a non mdt product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE INTEGRITY RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key6900363
Report Number9612164-2017-01336
Device Sequence Number10897252
Product Code NIQ
UDI-Device Identifier00643169037427
UDI-Public00643169037427
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P110013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/05/2018
Device Catalogue NumberRSINT22518UX
Device Lot Number0007926086
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2017
Initial Report FDA Received Date09/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-