• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CASHEL AVIATOR PLUS .014 5.0X15 142CM; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CASHEL AVIATOR PLUS .014 5.0X15 142CM; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4245015W
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/14/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that a 5x150 aviator plus balloon catheter was delivered to the lesion and inflated.However it ruptured at its nominal pressure.Therefore it was replaced with a same size new unknown balloon catheter and the procedure was completed successfully.There was no reported patient injury.The target lesion was unknown.The patient¿s vessel level of tortuosity and calcification was unknown.The rate of stenosis was unknown.The device prep normally.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.The product was removed intact (in one piece) from the patient.No other information was provided.The device was not returned for analysis.A device history record (dhr) review of lot 17273047 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause of the burst could not be determined.Based on the limited information available for review, vessel characteristics (although unknown) may have contributed to the reported event as calcification/resistant lesions may cause damage to a balloon.According to the instructions for use, which is not intended as a mitigation, ¿the rated burst pressure is based on the results of in vitro testing.At least (b)(4) of the balloons (with a (b)(4) confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Do not exceed the rated burst pressure recommended on the label.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the dhr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective/preventive action will be taken at this time.
 
Event or Problem Description
It was reported that a aviator plus balloon catheter was delivered to the lesion and inflated.However it ruptured at its nominal pressure.Therefore it was replaced with a same size new unknown balloon catheter and the procedure was completed successfully.There was no reported patient injury.The target lesion was unknown.The patient¿s vessel level of tortuosity and calcification was unknown.The rate of stenosis was unknown.The device prep normally.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.The product was removed intact (in one piece) from the patient.The product was clinically used and will not be returned for analysis.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVIATOR PLUS .014 5.0X15 142CM
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, tipperary
EI 
MDR Report Key6900813
Report Number9616099-2017-01471
Device Sequence Number16471988
Product Code LIT
UDI-Device Identifier20705032005811
UDI-Public20705032005811
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K071189
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 09/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2018
Device Model Number4245015W
Device Catalogue Number4245015W
Device Lot Number17273047
Was Device Available for Evaluation? No
Distributor Facility Aware Date09/14/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2017
Initial Report FDA Received Date09/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-