| Catalog Number |
6519-T-204 |
| Medical Device Problem Codes |
Device Operates Differently Than Expected (2913); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a tha, the trunnion sleeve did not fit completely (there was a gap) on the stem trunnion (citation stem).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: returned to manufacturer on; reason for no evaluation; an event regarding a size/fit issue mode involving a v40 sleeve was reported.The event was not confirmed.Method & results: device evaluation and results: visual inspection: the device was visually unremarkable.Dimensional inspection: a dimensional inspection was performed on the returned device.The device is dimensionally within specification.Medical records received and evaluation: not performed as no medical records were provided for review.Device history review: the device was manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other events for the lot referenced.Conclusions: a dimensional inspection was performed on the returned device.All dimensions were found to be within specification.No further investigation for this event is possible at this time.If devices and / or additional information become available, this investigation will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a tha, the trunnion sleeve did not fit completely (there was a gap) on the stem trunnion (citation stem).
|
| |
|