MEDTRONIC IRELAND ENDURANT ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Model Number |
ENLW1620C93E |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
09/11/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Other relevant device(s) are: catalog # enlw1613c124e, serial # (b)(4), use by date: 2013-08-02, manufactured date: 2011-08-05, and upn# (b)(4).Catalog # enew1313c82e, serial # (b)(4), use by date: 2012-11-22, manufactured date: 2010-11-24, and upn# (b)(4).A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant stent graft system was implanted in a patient for the endovascular treatment of a pre-operatively ruptured abdominal aortic aneurysm.It was reported that when the patient returned for follow up a dilated neck and distal seal zones were observed.It was reported that there were type ib endoleaks in both iliac arteries and that there was a proximal type i endoleak.The endoleaks appear to be related to the dilated proximal and distal zones.The physician explanted the stent grafts and treated surgically.The patient's creatinine has gone from 2.8 to 1.6.No additional clinical sequelae were reported and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received reported that there was no rupture prior to the implant or explant of the device.Device evaluation summary: the exact cause of the proximal type i and distal type i x2 endoleak could not be determined from the films provided.Earlier post-implant films were not available for review and comparison, and the post-implant films provided were non-contrast; therefore, no assessment of any endoleak could not be performed.However, it appears likely that the cause was due to disease progression.Pre-implant films revealed that the proximal neck diameter just below the renals measured ~25mm, was ~10mm in length, moderately angulated, and contained irregular thrombus with areas of calcification.Non-contrast films from 6 years post-implant showed that the bifurcate may have migrated into the sac.The aortic diameter at the level of the bifurcate proximal margin was 50mm, and the bifurcate angulated ~70 deg a-p relative to the distal aorta.There was also lack of distal limb radial apposition within both common iliac arteries.The rcia diameter had increased to 37mm and the lcia diameter had increased to 20mm.It appears most likely that disease progression was the primary cause of the endoleaks.Implanting within a short neck which contained thrombus, calcification, and was mod erately angulated may have also contributed to the type ia endoleak.If information is provided in the future, a supplemental report will be issued.
|
| |
|