| Catalog Number |
FFS000-01 |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: review of software logs were performed.A probe temperature was recorded in the hardware log files which showed evidence of possible unintended heating of the probe.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ablation procedure, there was unintended probe heating.There were no patient complications reported in relation to this event.If additional information is received, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated that there was no patient injury, and that the patient is seizure free.
|
| |
|
Search Alerts/Recalls
|