MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ALUMINA V40-FEMORAL HEAD 32MM, +4MM NK; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ALUMINA V40-FEMORAL HEAD 32MM, +4MM NK; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 6565-0-232
Device Problems Break (1069); Noise, Audible (3273)
Patient Problem Injury (2348)
Event Date 09/01/2017
Event Type  Injury  
Manufacturer Narrative
Review of the product history records indicates the products were manufactured and accepted into final stock with no reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event Description
It was reported that surgeon performed a revision of patient's hip (side unreported) due to an alumina head shattering.Patient reported to surgeon there was an audible clicking noise in the joint and when lifting weights, heard a loud pop.Rep reported that x-rays showed that the head had shattered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALUMINA V40-FEMORAL HEAD 32MM, +4MM NK
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
Manufacturer Contact
timothy rice
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6904171
MDR Text Key87767163
Report Number0002249697-2017-02889
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial
Report Date 09/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/23/2020
Device Catalogue Number6565-0-232
Device Lot Number52287702
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age51 YR
Patient Weight106
-
-