| Model Number |
CS 3805-075-050 |
| Medical Device Problem Codes |
Break (1069); Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017); Device Handling Problem (3265)
|
| Health Effect - Clinical Codes |
Pain (1994); Discomfort (2330)
|
| Date of Event |
09/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Ulrich medical gmbh (manufacturer) is submitting this report for both ulrich medical gmbh and ulrich medical usa (importer) exemption # e2014011.
|
| |
|
Event or Problem Description
|
|
Surgeon was inserting screw using a powered screwdriver and screwed it too far into the bone.The screw head broke off from the tulip and the broken pieces were removed and replaced.This caused a delay in surgery and the patient reported post-operative pain and discomfort that the surgeon attributed to this event.
|
| |
|
Search Alerts/Recalls
|