Model Number ZXR00
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implant and explant dates: if implanted; or explanted give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that after the intraocular lens (iol) was implanted, it appeared to have a mark or a scratch.The iol was cut in pieces to remove it from the eye, and a new iol (same model and diopter size) was implanted.The patient is fine.There was no incision enlargement, no serious patient injury, no sutures, and no vitrectomy.No additional information was provided to abbott medical optics.
 
Additional Manufacturer Narrative
Device available for evaluation? yes, returned to manufacturer on 9/28/2017.Device evaluation: visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens, additional analysis is not possible.The reported issue was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The dfu provide the customer with proper usage instructions and guidelines.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6906498
Report Number9614546-2017-00940
Device Sequence Number18967275
Product Code POE
UDI-Device Identifier05050474591660
UDI-Public(01)05050474591660(17)220331
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date03/31/2022
Device Model NumberZXR00
Device Catalogue NumberZXR00U0110
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2017
Initial Date Received by Manufacturer 09/06/2017
Supplement Date Received by Manufacturer10/13/2017
Initial Report FDA Received Date09/29/2017
Supplement Report FDA Received Date11/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-