| Model Number |
OS900041-NS |
| Medical Device Problem Codes |
Corroded (1131); Material Erosion (1214); Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/30/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complaint was received regarding a residue found inside the cannulation of the drills.This residue was identified as rust.A series of tests were conducted to rule out and identify the root cause however it is yet to be determined.In the meantime a recall will be initiated to remove the products from the field while ortho solutions continues to investigate this complaint.Material chemical composition test and device history record review showed the product was made to approved specification.Further test were conducted on identification of residue via edx.Tests are also being conducted to re-check the suggested drying time for steam sterilisation validation.The reported temperature and timeline was already validated for this system via independent laboratory.Further report will be submitted when recall is initiated and/or root cause is identified.
|
| |
|
Event or Problem Description
|
|
Brown residue was found inside the cannulation of the drills.It was found during cleaning and was not used on the patient.Ortho solutions have since conducted a series of test and the resiude was identified as early stage rust.Currently root case being investigated.
|
| |
|