Event or Problem Description
|
|
According to the reporter, during a laparoscopic hernia repair, the needle broke in the cavity and couldn't be retrieved right away.The doctor had to perform x-rays to find the needle and it was eventually retrieved.Surgery time was extended over 30 minutes and there was very little blood loss.The patient is currently fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of two device.Needle one pieces were broken at the suture swage.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the bent or broken needle may occur when the needle is not loaded properly or when the needle is forced into an obstacle.This condition may also occur if excess pressure is exerted on the needle or the attached suture while the jaws are in the open position.In these situations, the needle may flex and ultimately break.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|