Findings: pump received with the operating currents within spec.And passed the self test, unexpected alarm error test, rewind, basic occlusion test, occlusion test, prime test, excessive no delivery test, displacement test, and the dat test.Pump was programmed with multiple boluses and monitored.All boluses delivered properly and were listed in the bolus history screen.Pump then was programmed with multiple basal profiles and monitored.All basal profiles delivered their indicated amounts and were verified in the daily total screen.The units left at pump display matched properly the units left at test reservoir.No delivery anomaly, bolus anomaly or basal anomaly noted during testing.Pump received with cracked minor scratched lcd window and scratched reservoir tube window.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.
|