MAUDE Adverse Event Report: DATASCOPE FAIRFIELD YAMATO PLUS-R 7.5FR. 30CC IAB - JAPAN; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE FAIRFIELD YAMATO PLUS-R 7.5FR. 30CC IAB - JAPAN; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0549-01
Device Problems Backflow (1064); Leak/Splash (1354); Particulates (1451)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/11/2017
Event Type  malfunction  
Manufacturer Narrative
The product has been returned to the manufacturer, but is pending investigation.Once the investigation is completed a supplemental report with our findings will be submitted.(b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy a balloon leak and a blood back into the device was suspected.Removed iab and discontinued therapy.No patient injury was reported.Additionally, it was reported that there were white particles like spots on the tube, unknown if it was internal or external of the tube.Also unknown it was caused by calcification.
 
Manufacturer Narrative
The product was returned with the membrane completely unfolded and blood on the interior and exterior of the catheter and the membrane.The returned sheath was over the catheter but it was not a maquet product.The extension tubing was also returned.Dried blood was found occluding the inner lumen.The occlusion was unable to be cleared.An underwater leak test of the balloon, catheter and y-fitting was performed and one leak was detected on the membrane approximately 1.8cm from the rear seal measuring 0.02cm in length.The evaluation confirmed the reported problem.The reported problem was most likely triggered by a leak which was found on the membrane.Under magnification, a whitish patch was observed around the leak.This whitish patch is the typical appearance of an abrasion mark which is caused by calcified plaque in the aorta.A device and lot history record review was completed for the reported product.No non-conformances were found that are considered to be related to the event.(b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy a balloon leak and a blood back into the device was suspected.Removed iab and discontinued therapy.No patient injury was reported.Additionally, it was reported that there were white particles like spots on the tube, unknown if it was internal or external of the tube.Also unknown it was caused by calcification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
YAMATO PLUS-R 7.5FR. 30CC IAB - JAPAN
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer Contact
15 law drive
fairfield, NJ 07004
MDR Report Key6909191
MDR Text Key88120154
Report Number2248146-2017-00429
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date09/21/2019
Device Catalogue Number0684-00-0549-01
Device Lot Number3000036797
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2017
Is the Reporter a Health Professional? No
Device AgeYR
Initial Date Manufacturer Received 09/12/2017
Initial Date FDA Received10/02/2017
Supplement Dates Manufacturer Received10/26/2017
Supplement Dates FDA Received11/09/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-