BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM09100
Medical Device Problem Codes Positioning Failure (1158); Fracture (1260); Material Perforation (2205)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during deployment, using the pin and pull process, the endovascular stent graft allegedly only partially deployed as there was no more room between the catheter and the handle to continue with the pin & pull action in order to fully deploy the stent graft.Therefore, the delivery system was removed with the partial deployed stent without incident.Reportedly, another delivery system was used in order to successfully complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing related cause was performed.The lot history records were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.There was no manufacturing anomalies or changes which may have caused or contributed to the reported event.No additional complaints have been reported for this lot number previously.Investigation summary: on the basis of the returned stent graft delivery system, a deployment failure was confirmed.The outer sheath was found to be elongated and fractured, which indicated that increased friction affected the delivery system during attempt of stent graft deployment.Furthermore, the outer sheath was found perforated by stent graft strut, which made a stent graft deployment impossible.As a result of the investigation performed the complaint was confirmed.However, based on the available information and the evaluation of the returned sample, a definite root cause for the reported event could not be determined.Labeling review: in reviewing the labeling supplied with this product it was found that the instructions for use (ifu) sufficiently address the potential risk.The ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." regarding the anatomy of the placement site the ifu states: "the safety and effectiveness of the device when placed across a tight bend including the terminal cephalic arch or across the elbow joint has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." regarding the use of accessories the ifu states: "materials required for the fluency® plus endovascular stent graft procedure: (.) introducer sheath with appropriate inner diameter (.)".Furthermore, the ifu states: "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline (¿).Flushing these lumens will also facilitate stent graft deployment." (b)(4).
 
Event or Problem Description
It was reported that during deployment, using the pin and pull process, the endovascular stent graft allegedly only partially deployed as there was no more room between the catheter and the handle to continue with the pin & pull action in order to fully deploy the stent graft.Therefore, the delivery system was removed with the partial deployed stent without incident.Reportedly, another delivery system was used in order to successfully complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6909466
Report Number9681442-2017-00276
Device Sequence Number699463
Product Code PFV
UDI-Device Identifier04049519008561
UDI-Public(01)04049519008561
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/19/2018
Device Model NumberFEM09100
Device Catalogue NumberFEM09100
Device Lot NumberANZF2102
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2017
Initial Date Received by Manufacturer 09/05/2017
Supplement Date Received by Manufacturer12/14/2017
Initial Report FDA Received Date10/03/2017
Supplement Report FDA Received Date01/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age39 YR
-
-