• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CONQUEST 40 PTA BALLOON DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CONQUEST 40 PTA BALLOON DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number CQF5084
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Intimal Dissection (1333)
Date of Event 08/24/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the proximal basilic vein, the pta balloon allegedly ruptured on the first inflation attempt.The pta balloon was exchanged over the guidewire for another that was used to complete the procedure.Reportedly, a small vessel dissection occurred post balloon rupture.Patient was reported to be asymptomatic post procedure.
 
Additional Manufacturer Narrative
The device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Conclusion: the device was returned for evaluation.A visual inspection did not find any rupture related defects with the balloon.However, an inflation attempt was made and a longitudinal rupture was identified on the barrel of the balloon.Therefore, the investigation is confirmed for a longitudinal rupture.The definitive root cause for the identified rupture could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.Labeling review: the current ifu (instructions for use) states: warnings: do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.(b)(4).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the proximal basilic vein, the pta balloon allegedly ruptured on the first inflation attempt.The pta balloon was exchanged over the guidewire for another that was used to complete the procedure.Reportedly, a small vessel dissection occurred post balloon rupture.Patient was reported to be asymptomatic post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40 PTA BALLOON DILATATION CATHETER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6911881
Report Number2020394-2017-01267
Device Sequence Number125218
Product Code LIT
UDI-Device Identifier00801741060267
UDI-Public(01)00801741060267
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K120660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Model NumberCQF5084
Device Catalogue NumberCQF5084
Device Lot NumberREBU0571
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2017
Initial Date Received by Manufacturer 09/06/2017
Supplement Date Received by Manufacturer10/19/2017
Initial Report FDA Received Date10/03/2017
Supplement Report FDA Received Date11/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient Weight162
-
-