• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES SPINE MOUNTAINEER; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES SPINE MOUNTAINEER; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION Back to Search Results
Model Number 188319320
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Injury (2348); Device Embedded In Tissue or Plaque (3165)
Date of Event 09/05/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A complaint investigation will be performed.The complaint product is not available for the investigation.A supplemental report is not anticipated unless the results of the complaint investigation identify a corrective action or additional relevant information.Should the product become available, a physical evaluation will be conducted and a supplemental report filed with the results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Not returned for evaluation.
 
Event or Problem Description
On (b)(6) 2015, surgery for cervical spine trauma was performed using the mountaineer system.The fixed area was c4 - c6.Because of better bone growth, on (b)(6) 2017, removal surgery was performed due to synostosis.During the surgery, the surgeon took the following procedures.When he removed two set screws on the left side, it was confirmed that the head part of the screw (part #: 188319320, lot#: apfcwp) at c6 was loosened.Then, as he removed the rod, he noticed that the neck of this screw was broken.He tried to remove this screw shaft, however, he could not do that.Finally, he closed an incision with the broken screw shaft left in the bone.The surgery was completed without any delay.The surgeon comments that it does not matter if the broken screw shaft is left in the bone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOUNTAINEER
Common Device Name
ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION
Manufacturer (Section D)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
MDR Report Key6912134
Report Number1526439-2017-10796
Device Sequence Number10920888
Product Code NKG
UDI-Device Identifier10705034157658
UDI-Public(01)10705034157658
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K041203
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number188319320
Device Catalogue Number188319320
Device Lot NumberAPFCWP
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2017
Initial Report FDA Received Date10/03/2017
Patient Sequence Number1
Patient Age34 YR
-
-