| Model Number |
315-0A-714 |
| Health Effect - Clinical Codes |
Fall (1848); Rupture (2208)
|
| Date of Event |
09/04/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Second revision surgery - due to the patient having ruptured her patella tendon.She had been revised 2 days prior by the surgeon due to a post-op fall.He went back into fix that issue.When he did he also swapped the poly for a larger size.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for this revision surgery was the patient had a post operative fall and ruptured the patella tendon.The in-vivo length of patient service for the implant was 2 days.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient had a post-op fall and ruptured the patella tendon.The surgeon repaired the tendon and exchanged the tibia insert.This event occurred after the first revision.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
|
| |
|
Search Alerts/Recalls
|