MAUDE Adverse Event Report: COSMAN MEDICAL NITINOL TC ELECTRODE; ELECTRODES
FDA Home
Medical Devices
Databases
-
510(k)
|
DeNovo
|
Registration & Listing
|
Adverse Events
|
Recalls
|
PMA
|
HDE
|
Classification
|
Standards
CFR Title 21
|
Radiation-Emitting Products
|
X-Ray Assembler
|
Medsun Reports
|
CLIA
|
TPLC
COSMAN MEDICAL NITINOL TC ELECTRODE; ELECTRODES
Back to Search Results
Model Number
CSK-TC10-3M
Medical Device Problem Code
Break (1069)
Health Effect - Clinical Code
No Known Impact Or Consequence To Patient (2692)
Date of Event
09/08/2017
Type of Reportable Event
Malfunction
Event or Problem Description
The device analysis performed on the returned electrode showed that the shaft was broken off at the hub due to external mechanical force.
Search Alerts/Recalls
New Search
|
Submit an Adverse Event Report
Brand Name
NITINOL TC ELECTRODE
Common Device Name
ELECTRODES
Manufacturer
(Section D)
COSMAN MEDICAL
22 terry ave
burlington MA 01803
MDR Report Key
6912325
Report Number
3006630150-2017-03918
Device Sequence Number
117120
Product Code
GXI
UDI-Device Identifier
00813250011524
UDI-Public
00813250011524
Combination Product (Y/N)
N
Initial Reporter State
FL
Initial Reporter Country
US
PMA/510(K) Number
K050084
Number of Events Summarized
1
Summary Report (Y/N)
N
Reporter Type
Manufacturer
Report Source
company representative,health
Initial Reporter Occupation
Health Professional
Type of Report
Initial
Report Date
(Section B)
09/08/2017
1
Device was Involved in the Event
1
Patient was Involved in the Event
Is this an Adverse Event Report?
No
Is this a Product Problem Report?
Yes
Operator of Device
Physician
Device Model Number
CSK-TC10-3M
Device Lot Number
062816
Was Device Available for Evaluation?
Device Returned to Manufacturer
Date Returned to Manufacturer
08/24/2017
Type of Report(Section G)
Thirty-Day
Initial Date Received by Manufacturer
09/08/2017
Initial Report FDA Received Date
10/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number
1
-
-