• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C0530
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/08/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report numbers: 3005168196-2017-01712.3005168196-2017-01713.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the portal vein using ruby coils.During the procedure, while attempting to advance a ruby coil through a lantern delivery microcatheter (lantern), the physician experienced resistance after advancing the coil about half way out of the lantern and the coil would not advance further; therefore, it was removed.The physician then attempted to advance a new ruby coil through the lantern; however, resistance was encountered again and the coil would not completely advance out of the lantern.Therefore, it was removed.The physician then removed the lantern to flush it; however, the lantern was accidentally dropped on the floor.Therefore, the procedure was completed using two additional ruby coils and a new lantern.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note the device is no longer available for return as mentioned in the initial mdr; therefore, the mdr was updated accordingly.The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The device was lost in transit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6912500
Report Number3005168196-2017-01711
Device Sequence Number13628242
Product Code HCG
UDI-Device Identifier00814548012919
UDI-Public00814548012919
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/31/2024
Device Catalogue NumberRBY2C0530
Device Lot NumberF71445
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/08/2017
Supplement Date Received by Manufacturer04/10/2018
Initial Report FDA Received Date10/03/2017
Supplement Report FDA Received Date04/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age49 YR
-
-