The returned device did not match initial reported condition.Additional information has been requested based upon the condition of the returned device, however no response has been received.Further investigation is in progress.When completed, a supplemental report will be submitted.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
As received, the pipeline flex pushwire was observed sepa rated into two segments (distal and proximal).The separation was not originally reported.The pipeline braid was returned detached from the pushwire and found fully opened.The pipeline braid was found fraying at the both ends.Additionally, the distal segment of the pipeline flex pushwire including the tip coil and distal protective sleeves (dps) were missing.The pushwire bent at the proximal end.The distal section of the proximal segment of the pipeline flex pushwire was sent out for scanning electron microscopy (sem) analysis.Follow up has been done in regard to the missing distal segment; however, no responses have been received.Based on the sem analysis, the failure mechanism for the separation was wire failure via torsional overload failure.It is possible that the reported ¿resistance¿ may have contributed to damages observed on the pipeline flex; it appears excessive force was used (pushing and pulling).Per our instructions for use (ifu): ¿it is recommended to use a heparinized saline drip to continuously flush micro catheter during pipeline flex embolization device use.If high forces or excessive friction is encountered during delivery, discontinue delivery of the device and identify the cause of the resistance, remove device and micro catheter simultaneously.Advancement of the pipeline flex embolization device against resistance may result in damage or patient injury.¿ no evidence was found to suggest that the devices failed to meet specifications; therefore, manufacturing has been ruled out as a potential cause.All products are 100% inspected for damage and irregularities during manufacture.If information is provided in the future, a supplemental report will be issued.
|