BARD PERIPHERAL VASCULAR, INC. CONQUEST 40 PTA BALLOON DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number CQF7564
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a left upper arm fistula, the health care provider allegedly had difficulties deflating the pta balloon.Reportedly, the original pta balloon was used to complete the procedure and subsequently was successfully removed through the sheath at the conclusion of the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Conclusion: the device was returned for evaluation.A visual inspection found no deflation related defects with the device.The returned sample was inflated and deflated without any issues within the specified time constraints.Therefore, the investigation is unconfirmed for deflation issues as the balloon functioned as intended during functional testing.The definitive root cause could not be determined based upon available information.It is unknown whether patient and/or procedural issues contributed to the event.It should be noted that per the reported event details, it was stated that the device was inflated with a syringe.The ifu (instructions for use) states that the use of a pressure monitoring device is recommended to prevent over pressurization.It is unknown whether the balloon/catheter was overpressurized or the use of a syringe contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.When the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Precautions: if resistance is felt during post procedure withdrawal of the catheter through the introducer sheath, determine if contrast is trapped in the balloon with fluoroscopy.If contrast is present, push the balloon out of the sheath and then completely evacuate the contrast before proceeding to withdraw the balloon.If resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/introducer sheath as a single unit.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a left upper arm fistula, the health care provider allegedly had difficulties deflating the pta balloon.Reportedly, the original pta balloon was used to complete the procedure and subsequently was successfully removed through the sheath at the conclusion of the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40 PTA BALLOON DILATATION CATHETER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6915306
Report Number2020394-2017-01278
Device Sequence Number1312870
Product Code LIT
UDI-Device Identifier00801741060441
UDI-Public(01)00801741060441
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K120660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2019
Device Model NumberCQF7564
Device Catalogue NumberCQF7564
Device Lot NumberREAV1592
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2017
Initial Date Received by Manufacturer 09/07/2017
Supplement Date Received by Manufacturer10/17/2017
Initial Report FDA Received Date10/05/2017
Supplement Report FDA Received Date10/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
Patient Weight83
-
-