| Catalog Number |
AB35W06040080 |
| Medical Device Problem Codes |
Burst Container or Vessel (1074); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Patient Problem/Medical Problem (2688); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
09/09/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician was attempting to use an evercross pta balloon catheter in a patient when the balloon burst.The balloon was removed from the patient but a portion of the balloon remained in the patient.A stent was deployed to jail the balloon fragment against the wall of the artery.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary the evercross dilatation catheter was returned for evaluation.A portion of 6mm by 40mm balloon chamber material, inner guidewire lumen approximately 40mm including the distal tip, and two balloon chamber radiopaque marker bands were not returned.No ancillary devices or cine images from the procedure were received.The balloon chamber material separation face appears to be a radial tear within the proximal balloon chamber cone and the inner guidewire lumen separation exhibits a tensile separation at the proximal end of the proximal radiopaque marker band.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Event description updated with additional information inflation atm was reported as 8 atm which is below the rbp for this device (14 atm) a 6fr sheath and a 0.035 guidewire was used in the procedure no embolic protection was used the physician used a new balloon and it worked with no issues reported.If information is provided in the future, a supplemental report will be issued.
|
| |
|