Medtronic received report that after the device was successfully delivered, it was realized that the device separated at the proximal bumper.The patient was initially asymptomatic, however, a few hours later was reported to have stroke like symptoms.The patient was on antiplatelet medication.The anatomy was moderate to severe.A continuous flush was used during the procedure.It is believed that the separated pushwire may be cause of the post procedure symptoms.
|
The proximal segment of the flow diverter¿s pushwire was returned for evaluation.The braid and the broken distal segment of the pushwire were not returned, as the braid was implanted with the broken distal pushwire.The micro catheter was not returned for evaluation, as it was discarded.The pushwire was found to be separated at the distal hypotube.The hypotube was found to be severely stretched.The pushwire was found to be kinked/ bent near the proximal end.No other anomalies were observed.The broken end of the pushwire was then sent out for scanning electron microscopy (sem) analysis.Based on the analysis findings, the event details and the sem analysis; the report of pushwire break at distal hypotube was confirmed, as the received pushwire was found to be separated at the distal hypotube.The fracture surface features of the broken end are consistent with tensile overload failure.From the damages seen on the proximal wire (kinking/bending) and hypotube (severely stretching); it appears that excessive force was used.It is possible the severe vessel tortuosity may contribute to the excessive force used during procedure; subsequently causing the distal hypotube to become separated.Per our instructions for use: ¿discontinue delivery of the device if high force or excessive friction is encountered.Identify the cause of the resistance and remove device and microcatheter simultaneously.Advancement of the ped against resistance may result in device damage or patient injury.Never advance or withdraw an intraluminal device against resistance until the cause of resistance is determined by fluoroscopy.If the cause cannot be determined, withdraw the catheter.Movement of the micro catheter against resistance may result in damage to the micro catheter, or the vessel.Do not use in patients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿ all products are 100% inspected for damage and irregularities during manufacture.Updated with udi number.If information is provided in the future, a supplemental report will be issued.
|