Model Number 16-02-80
Device Problems Failure to Calibrate (2440); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 09/10/2017
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported failure and noted an error code on patient circuit 2.The level sensor was checked and re-positioned.Through further troubleshooting, the fault was traced to the cpu distributor connector, which was out of alignment.The alignment was corrected and a functional test was performed without further issues.The device was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Device not returned.
 
Event Description
Livanova (b)(4) received a report that the temperature calibration of the heater-cooler system 3t could not be carried out during maintenance, as the level sensor indicated that the tank was empty when it was not.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6916090
MDR Text Key88235674
Report Number9611109-2017-00776
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 10/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-