| Model Number |
ZXT225 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Visual Disturbances (2140); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is during 2017.(b)(4).All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxt225 21.0 diopter intraocular lens (iol) was implanted in the right eye (od) on (b)(6) 2017.It was later explanted on (b)(6) 2017 due to patient complaints about being intolerant to starbursts experienced.There was no incision enlargement, vitrectomy, or sutures used.There was no post-operative patient injury reported.The lens was replaced with a the same diopter of 21.0, but different model zct150.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation?: yes, returned to manufacturer on 10/05/2017.Device returned to manufacturer? yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|