Model Number 1032804
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Death (1802)
Date of Event 09/08/2017
Type of Reportable Event Death
Event or Problem Description
The manufacturer received information alleging a patient expired after becoming disconnected from a ventilator.The reported day of the event was (b)(6) 2017.According to the reporter of the event, the patient's wife checked on him at approximately 15:00:00 local time on (b)(6) 2017 and then returned to work.A relative visited the house and found the patient disconnected from the ventilator.The relative reconnected the patient and attempted to contact the wife.The patient was taken to a hospital where he expired.The ventilator was returned to the manufacturer for evaluation.The device passed all testing and was found to operate and alarm as designed.The ventilator's downloaded event logs were reviewed.The device maintains the event logs in non-volatile memory.An event, as logged by the device, includes every keypress, alarm or failure of the device to remain within specifications.There were no events logged related to device performance or device malfunctions.If a device malfunction occurred, the event logs would have contained entries related to the malfunction.On the reported day of the event, (b)(6) 2017, the ventilator began alarming for a low peak expiratory pressure condition at 17:45:21 local time followed by low peak inspiratory pressure and patient circuit disconnect alarms.The ventilator alarmed for 3458 seconds (approximately 57.6 minutes).This issue appears to have been corrected.At 18:59:43 on (b)(6) 2017, the ventilator began alarming for low peak expiratory pressure and low peak inspiratory pressure followed by a patient circuit disconnect alarm.The ventilator alarmed for 463 seconds (approximately 7.7 minutes) and the device was manually powered off at 19:07:26 local time.The ventilator was not powered on again until 23:29:22 on (b)(6) 2017.The manufacturer concludes the ventilator operated and alarmed as designed and did not cause or contribute to the reported event.The ventilator alarmed to alert the caregiver of the patient circuit disconnect which was not acknowledged in a timely manner.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 200
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS INC.
1001 murry ridge lane
murrysville PA 15668
MDR Report Key6916787
Report Number2518422-2017-02259
Device Sequence Number1444283
Product Code CBK
UDI-Device Identifier00606959033863
UDI-Public00606959033863
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093416
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 09/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Patient Family Member or Friend
Device Model Number1032804
Device Catalogue Number1032804
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/24/2017
Initial Report FDA Received Date10/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age64 YR
-
-