This report was initially submitted following a report of discrepant results associated with api® nh (reference 10400).The customer stated the expected identification as neisseria sicca; however, the result obtained was neisseria gonorrhea 98% (good id).Repeat testing confirmed the misidentification.An internal biomérieux investigation was performed.Due to the expiration (28sep2017) of the incriminated batch (lot 1005423530), testing was not possible; however, the strain, atcc 9913 neisseria sicca, was tested via api nh strips.The results were compliant with a "very good identification" to neisseria spp at 99.9%.Analysis of the batch record, including bacteriological tests, indicates all criteria for release were compliant.According to the package insert rev 07487p-xl-2016/12 (current version), the knowledge base of the strip is limited to neisseria spp.Complementary tests are necessary to identify the genus to neisseria sicca.Conclusion: due to the expiration of the impacted api nh strips, testing couldn't be performed in the biomérieux qc laboratory.However, a previous study on strain atcc 9913 neisseria sicca at biomérieux showed this strain is correctly identified with the api nh strips.
|