• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COSMAN MEDICAL CSK-TC ELECTRODE; ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COSMAN MEDICAL CSK-TC ELECTRODE; ELECTRODES Back to Search Results
Model Number CSK-TC10-3M
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/14/2017
Type of Reportable Event Malfunction
Event or Problem Description
Device analysis on the returned electrode revealed traces of potting epoxy on the hub.The epoxy leaked through the shaft opening in its liquid state during the manufacturing process, and was not removed after curing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSK-TC ELECTRODE
Common Device Name
ELECTRODES
Manufacturer (Section D)
COSMAN MEDICAL
22 terry ave
burlington MA 01803
MDR Report Key6917633
Report Number3006630150-2017-03609
Device Sequence Number13781177
Product Code GXI
UDI-Device Identifier00813250011524
UDI-Public00813250011524
Combination Product (Y/N)N
Initial Reporter StateMO
PMA/510(K) Number
K050084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 09/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model NumberCSK-TC10-3M
Device Lot NumberL538
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2017
Initial Report FDA Received Date10/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-